Pico Technology Oscilloscope Probe, Probe Type: Active, Differential 100MHz 700V

產品概覽和技術數據資料表
法例與合規
相容
COO (Country of Origin): TW
產品詳細資訊

Pico 100MHz 700V Differential Probe TA043, 10:1/100:1

Pico TA043 is an active differential probe suitable for high common-mode voltage measurement applications up to ±700V (DC + peak AC). It is capable of acquiring higher speed signals of up to 100MHz. This differential probe extends the functionality of standard single-ended input oscilloscopes to allow a safe and accurate method of making high-voltage differential measurements. Applications include making safe measurements in power circuit applications and acquisition of low speed balanced differential signals found in serial communications buses.

Bandwidth DC to 100MHz (–3 dB)
Attenuation 10:1/100:1
Rise time 3.5ns
Accuracy ±2%
Input impedance 4MΩ ll 7pF each side to ground
Input voltage differential range (10:1) ±70V (DC + peak AC) or 70VRMS
Input voltage differential range (100:1) ±700V (DC + peak AC) or 500VRMS
Common mode range ±700V (DC + peak AC) or 500VRMS (10:1 and 100:1)
Absolute maximum voltage (differential or common mode) 1000VRMS CAT III (10;1 & 100:1)
Swing (into 50 kΩ load) ±7V
Offset (typical) <±5mV
Noise (typical) 0.9mVRMS
Source impedance (typical) 50Ω
CMRR (typical) 85dB @ 50Hz; 55dB @ 1MHz
Ambient operating temperature -10 to +40°C
Ambient storage temperature -30 to +70°C
Power requirement 4xAA cells or 6VDC/200mA mains adaptor or regulated 9VDC/120mA mains adaptor
Length of input leads 30cm
Length of BNC lead 90cm
Weight 500g
Dimensions 202 x 83 x 38 mm

Pico Technology Oscilloscope Probe

The Pico Technology TA043 oscilloscope probe is an active differential probe, suitable for high common-mode voltage measurement applications up to ±7000V. It can be used with a signal speed of up to 100MHz. This differential probe extends the functionality of standard single-ended input oscilloscopes to allow a safe and accurate method of making high voltage differential measurements.

Features and Benefits

• Bandwidth is DC to 100MHz (–3dB)
• Attenuation ratio is 10:1/100:1
• Accuracy is ±2%
• Input impedance is 4Mohm ll7pF each side to ground
• Differential range is (10:1) ±70V (DC + peak AC) or 70VRMS
• Maximum differential is (100:1) ±700V (DC + peak AC) or 500VRMS
• Common mode range is ±700V (DC + peak AC) or 500VRMS (10:1 & 100:1)
• Absolute maximum voltage is ±1000VRMS CATIII 10:1 & 100:1)
• Swing is ±7V into 50Kohm load
• Offset (typical) is <±5mV
• Noise (typical) is 0.9mVRMS
• Source impedance is 50
• CMRR (typical) is 85dB at 50Hz and 55dB at 1MHz
• Ambient operating temperature ranges between -10°C and 40°C
• Power requirement 4xAA cells or 6VDC/200mA mains adaptor or regulated 9VDC/120mA mains adaptor

Applications

• Safe measurement in power circuit
• Acquisition of low speed balanced differential signals found in serial communication buses
• Transients in switching power supplies, inverters and motor drives, electronic ballasts and CRT displays
• Signals in the milliV range from sensors and other instrumentation

Certifications

• IEC-1010.1
• IEC-1010.2-031 CAT III
• Pollution Degree 2.

規格
Attribute Value
Probe Type Active, Differential
Attenuation 1:10, 1:100
Bandwidth 100MHz
Maximum Common Mode Input Voltage 700V
Number of Probes 2
Safety Category Level CAT III
Safety Category Voltage 500 Vrms, 700 V ac/dc
Safety Category CAT III 700 V ac/dc, 500 Vrms
Rise Time 3.5ns
14 現貨庫存,可於3工作日發貨。
單價 個
HK$ 5,046.98
(不含稅)
單位
Per unit
1 +
HK$5,046.98